DA3300
Anesthesia Machine
Gas Supply:Gas Source: O2,or O2&Air(Option), or O2&N2O(Option); Gas Source Pressure Gauge: O2,or O2&Air(Option), or O2&N2O(Option); Pressure of Pipeline Inlet: O2,280kPa~600kPa/41Psi~87Psi, less than 180L/min;N2O(Option),280kPa~600kPa/41Psi~87Psi,less than180L/min; Air(Option),280kPa~600kPa/41Psi~87Psi, less than 180L/min;Security Valve: release pressure less than 10 kpa;O2 Lost Alarm(Option):Sound alarm, alimumun cylinder with whistle;N2O cut off valve protection(Option): when O2 is lost, N2O will be cut off.Flow Meter:2 tubes high precision flow meter,0~1L/min,1.1L/min~10L/min; or 0L/min~10L/min(O2&Air or O2&N2O);Precision: At 20 degree, atmospheric pressure is 101.3kPa, the precision of flow meter is level:2.5;O2&N2O Linkage Device(O2&N2O,Option):make insure O2 concentration is not lower than 25%;O2 Flush: 35L/min~75L/min;One Key Switch: Switch between rebreath and Non-Rebreath;ACGO Port: Non-Rebreath;Selectatec Mounting Bar: single place used for vaporizer, it is easy to replace;
Vaporizer;Anesthesia Vaporizer: one of Isoflurane/Halothane/Sevoflurane;Respiration System/CO2 Circle Absorber/Rebreath:Soda Lime Chamber: 800ml or 1600ml(Option); Pop Off Valve: with one key to close; APL valve(Option): 0.19~6kPa;Airway Pressure Gauge:-20-100cmH2O;
Anesthesia Ventilator
Drive gas: O2 or Air, Inlet pressure: 280-600 kPa. Max flow≤ 120 L / min;Safety Pressure: System pressure does not exceed 10 KPa;Display:10.2" TFT, with touch screen;Breath Mode:VCV,PCV,SIMV-VC,SIMV-PC,PSV,Manual/SPONT,SIMV-PRVC(option);Waveform:Pressure,Flow rate,Volume, Loop,PV loop, PF loop,FV loopO2 concentration Monitor:15%-100%(option);
Electric PEEP: OFF, 3 ~ 30 cmH2O;
Battery Type:Li-ion,work time more than 3 hours;
Parameters setting range:
Tidal volume:50 - 1500ml;
option :10 - 1500ml;
Respiratory rate:1 ~ 100 bpm;
Inspiratorytime:0.1~10.0s;
Respiratory rate:0 ~ 100 bpm;Percentage of inspiratory pause:0 to 60%; PEEP:OFF, 3 ~ 30 cmH2O;
Pressure Support:0 ~ 70 cmH2O; Pressure Control:5 ~ 70 cmH2O;
Flow trigger:0.5 ~ 20L/min;
Pressure Trigger:0 ~ 20 cmH2O;
PSV Insp Termination Level:5 ~ 80%;
Monitored parameters:Inspiratory tidal volume: 0 ~ 3000 mL;Expiratory tidal volume:0 ~ 3000 mL;Spontaneous minute ventilation:0 ~ 60 L / min;Minute ventilation:0 ~ 60 L / min;Respiratory ratio:4:1 to 1:10;
Spontaneous breathing frequency:0 ~ 100 bpm;
Compliance:0 ~ 300 mL/cmH2O;Respiratory ratio:30:1 to 1:150;Peak airway pressure:0 ~ 100 cmH2O;
Mean airway pressure:0 ~ 100 cmH2O;PEEP:0 ~ 100 cmH2O;Inspiratory plateau pressure:0 ~ 100 cmH2O;
FiO2(option):15 to 100%;
Alarm settings:
Tidal volume high/low alarm; Minute ventilation high/low alarm; Respiratory rate high/low alarm; Airway pressure high/low alarm; Continuous Positive Airway Pressure/Negative pressure alarm; Apnea alarm; Oxygen deficiency alarm; AC power failure alarm ,Low battery alarm.
Standard configuration:
Anesthesia machine, one unit X60 Anesthesia vaporizer, one set adult disposable breath circuit, one set pediatric disposable breath circuit, 0.5,1.2,3L airbag, O2 gas source pipeline, one set trolley.
Dimension&Weight:Trolley Stand: 55cm*56cm*153cm,40kg;
Package Information:
Main unit: 71cm*33cm*75cm,Gross weight:29kg
Anesthesia Vaporizer:33cm*22cm*33cm, Gross weight:7.5kg
Trolley:69cm*57cm*26cm,Gross weight:10kg.
Whole gross weight:46.5kg